Pacemaker Lead Perforation Leading to Cardiac Tamponade and Subsequent Anterior STEMI
Recommended Citation
Martinez-Ponce J, Ubysz O, Costa A, Ataya L, Bhangu S, Vanhecke T. Pacemaker Lead Perforation Leading to Cardiac Tamponade and Subsequent Anterior STEMI. JACC Case Rep. 2026;31(24):108324.
Document Type
Article
Publication Date
6-17-2026
Publication Title
JACC Case Rep
Abstract
BACKGROUND: Pacemaker lead perforation is an uncommon but potentially fatal complication of cardiac implantable electronic devices, particularly when hemopericardium and tamponade occur in patients requiring antithrombotic therapy.
CASE REPORT: A 74-year-old woman with atrial fibrillation on apixaban and a recent dual-chamber pacemaker implantation presented with dizziness, falling, and dysarthria; her brain computed tomography was negative. She developed syncope with a 6-second ventricular pause and enlarging pericardial effusion and an emergent surgical pericardial window with drainage of 350 mL of hemorrhagic fluid, confirming right ventricular lead perforation left in situ. Shortly after stabilization, she developed anterior ST-segment elevation myocardial infarction because of proximal left anterior descending artery occlusion that was treated with percutaneous coronary intervention. Echocardiogram showed a left ventricular ejection fraction of 35% to 40% without recurrent effusion.
CONCLUSIONS: Early recognition, prompt drainage, and careful balancing of bleeding and thrombosis risks are essential; multidisciplinary management enables favorable outcomes.
TAKE-HOME MESSAGE: Maintain suspicion for lead perforation despite a normal device interrogation, and act rapidly in the event of tamponade while coordinating multidisciplinary care.
PubMed ID
42132721
Volume
31
Issue
24
First Page
108324
Last Page
108324