Time-dependent risk reduction of ventricular tachyarrhythmias in cardiac resynchronization therapy patients: a MADIT-RIT sub-study
Recommended Citation
Stockburger M, Moss AJ, Olshansky B, Klein H, McNitt S, Schuger C, Daubert JP, Goldenberg I, Ruwald AC, Merkely B, Zareba W, Kutyifa V. Time-dependent risk reduction of ventricular tachyarrhythmias in cardiac resynchronization therapy patients: a MADIT-RIT sub-study. Europace. 2015;17(7):1085-91.
Document Type
Article
Publication Date
7-1-2015
Publication Title
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
Keywords
Age Distribution, Aged, Aged, 80 and over, Cardiac Resynchronization Therapy, Defibrillators, Implantable, Disease-Free Survival, Female, Humans, Male, Middle Aged, Survival Analysis, Tachycardia, Ventricular, Treatment Outcome, United States, Ventricular Fibrillation
Abstract
AIMS: Data on the time-dependent benefit of cardiac resynchronization therapy with defibrillator (CRT-D) compared with a dual-chamber implantable cardioverter-defibrillator (ICD) to reduce death or ventricular tachycardia (VT) or ventricular fibrillation (VF) are limited. We aimed to evaluate the time-related risk of death or sustained VT or VF in patients receiving CRT-D vs. ICD in the MADIT-RIT trial.
METHODS AND RESULTS: Kaplan-Meier survival analyses and multivariate Cox regression models were utilized to compare the incidence and the risk of death or sustained VT/VF in the CRT-D and ICD subgroups by the elapsed time after device implantation (6 months). Of the ICD (n = 742) and CRT-D (n = 757) patients enrolled, the risk of death was lower in CRT-D vs. in ICD early after device implantation [hazard ratio (HR) = 0.42, 95% confidence interval (CI): 0.17-1.03, P = 0.058] and beyond 6 months of follow-up (HR = 0.39, 95% CI: 0.21-0.73, P = 0.004), with the 6-month interaction P = 0.899. The overall risk of sustained VT/VF was reduced in CRT-D vs. ICD patients (HR = 0.73, 95% CI: 0.52-1.03, P = 0.07). However, the risk was similar in the first 6 months (HR = 1.00, 95% CI: 0.62-1.62, P = 0.988), and a lower risk emerged 6 months after CRT-D implantation (HR = 0.58, 95% CI: 0.38-0.88, P = 0.011), with the 6-month interaction P = 0.059.
CONCLUSION: The reduced mortality risk of CRT-D compared with an ICD alone began early after device implantation and was sustained during long-term follow-up; the reduced risk for ventricular tachyarrhythmias did not emerge until 6 months after device implantation.
CLINICAL TRIAL REGISTRATION: http://clinicaltrials.gov/ct2/show/NCT00947310.
Medical Subject Headings
Age Distribution; Aged; Aged, 80 and over; Cardiac Resynchronization Therapy; Defibrillators, Implantable; Disease-Free Survival; Female; Humans; Male; Middle Aged; Survival Analysis; Tachycardia, Ventricular; Treatment Outcome; United States; Ventricular Fibrillation
PubMed ID
25745075
Volume
17
Issue
7
First Page
1085
Last Page
1091