Complications and failure modes of coronary embolic protection devices: Insights from the MAUDE database

Document Type

Article

Publication Date

4-20-2021

Publication Title

Catheterization and cardiovascular interventions

Abstract

BACKGROUND: There is limited data on complications associated with the use of coronary embolic protection devices (EPDs).

METHODS: We queried the Manufacturer and User Facility Device Experience database between November 2010 and November 2020 for reports on coronary EPDs: Spider FX (Medtronic, Minneapolis, MN) and Filterwire EZ (Boston Scientific, Natick, MA).

RESULTS: We retrieved 119 reports on coronary EPD failure (Spider FX n = 33 and Filterwire EZ n = 86), most of which (78.2%) occurred during saphenous vein graft interventions. The most common failure mode was inability to retrieve the EPD (49.6%), with the filter trapped against stent struts in 76.2% of the cases. Other device complications included filter fracture (28.6%), failure to cross (7.6%), failure to deploy (7.6%), and failure to recapture the filter (3.4%). Filter fracture (54.5 vs. 29.1%) and failure to recapture (9.1 vs. 2.1%) were more commonly reported, while failure to deploy the filter (0 vs. 10.5%) was less commonly reported with the Spider-FX.

CONCLUSIONS: The most common modes of failure of coronary EPDs are the failure of retrieval (49.6%), followed by the filter fracture (28.6%). When using EPDs, careful attention to the technique is essential to avoid failures and subsequent complications.