Can a Point-of-Care Troponin I Assay be as Good as a Central Laboratory Assay? A MIDAS Investigation.
Recommended Citation
Peacock WF, Diercks D, Birkhahn R, Singer AJ, Hollander JE, Nowak R, Safdar B, Miller CD, Peberdy M, Counselman F, Chandra A, Kosowsky J, Neuenschwander J, Schrock J, Lee-Lewandrowski E, Arnold W, and Nagurney J. Can a point-of-care troponin I assay be as good as a central laboratory assay? A MIDAS investigation. Ann Lab Med 2016; 36(5):405-412.
Document Type
Article
Publication Date
9-1-2016
Publication Title
Ann Lab Med
Abstract
BACKGROUND: We aimed to compare the diagnostic accuracy of the Alere Triage Cardio3 Tropinin I (TnI) assay (Alere, Inc., USA) and the PathFast cTnI-II (Mitsubishi Chemical Medience Corporation, Japan) against the central laboratory assay Singulex Erenna TnI assay (Singulex, USA).
METHODS: Using the Markers in the Diagnosis of Acute Coronary Syndromes (MIDAS) study population, we evaluated the ability of three different assays to identify patients with acute myocardial infarction (AMI). The MIDAS dataset, described elsewhere, is a prospective multicenter dataset of emergency department (ED) patients with suspected acute coronary syndrome (ACS) and a planned objective myocardial perfusion evaluation. Myocardial infarction (MI) was diagnosed by central adjudication.
RESULTS: The C-statistic with 95% confidence intervals (CI) for diagnosing MI by using a common population (n=241) was 0.95 (0.91-0.99), 0.95 (0.91-0.99), and 0.93 (0.89-0.97) for the Triage, Singulex, and PathFast assays, respectively. Of samples with detectable troponin, the absolute values had high Pearson (R(P)) and Spearman (R(S)) correlations and were R(P)=0.94 and R(S)=0.94 for Triage vs Singulex, R(P)=0.93 and R(S)=0.85 for Triage vs PathFast, and R(P)=0.89 and R(S)=0.73 for PathFast vs Singulex.
CONCLUSIONS: In a single comparative population of ED patients with suspected ACS, the Triage Cardio3 TnI, PathFast, and Singulex TnI assays provided similar diagnostic performance for MI.
Medical Subject Headings
Acute Coronary Syndrome; Biomarkers; Emergency Service, Hospital; Humans; Laboratories; Myocardial Infarction; Point-of-Care Systems; Prospective Studies; Reagent Kits, Diagnostic; Sensitivity and Specificity; Troponin I
PubMed ID
27374704
Volume
36
Issue
5
First Page
405
Last Page
412