AAPM Task Group 334: A guidance document to using radiotherapy immobilization devices and accessories in an MR environment
Recommended Citation
Hobson MA, Hu Y, Caldwell B, Cohen GN, Glide-Hurst C, Huang L, Jackson PD, Jang S, Langner U, Lee HJ, Levesque IR, Narayanan S, Park JC, Steffen J, Wu QJ, and Zhou Y. AAPM Task Group 334: A guidance document to using radiotherapy immobilization devices and accessories in an MR environment. Med Phys 2024.
Document Type
Article
Publication Date
6-1-2024
Publication Title
Medical physics
Abstract
Use of magnetic resonance (MR) imaging in radiation therapy has increased substantially in recent years as more radiotherapy centers are having MR simulators installed, requesting more time on clinical diagnostic MR systems, or even treating with combination MR linear accelerator (MR-linac) systems. With this increased use, to ensure the most accurate integration of images into radiotherapy (RT), RT immobilization devices and accessories must be able to be used safely in the MR environment and produce minimal perturbations. The determination of the safety profile and considerations often falls to the medical physicist or other support staff members who at a minimum should be a Level 2 personnel as per the ACR. The purpose of this guidance document will be to help guide the user in making determinations on MR Safety labeling (i.e., MR Safe, Conditional, or Unsafe) including standard testing, and verification of image quality, when using RT immobilization devices and accessories in an MR environment.
Medical Subject Headings
Magnetic Resonance Imaging; Humans; Immobilization; Radiotherapy, Image-Guided
PubMed ID
38648857
ePublication
ePub ahead of print
Volume
51
Issue
6
First Page
3822
Last Page
3849